In this double-blinded, randomized trial, authors measured 2-D, 3-D, and volumetric wear (in mm or mm2), along with wear rates (mm/year), presence or absence of osteolysis, and revision rates in 91 patients at specified time intervals, up to a minimum of 10 years. The following results corroborate the general findings from most other studies on this topic:
- The mean 3-D wear rate among patients with the XLPE acetabular liner was 0.03 mm/yr, versus 0.27 mm/yr among patients with conventional PE.
- Eight percent of patients in the XLPE group showed radiographic evidence of osteolysis, versus 38% of patients in the PE group.
- Patients with the conventional PE liner had a significantly higher revision rate (14.6%) than those with the XLPE liner (1.9%).
There were no significant between-group differences in clinical outcome scores, including the Oxford Hip Score and SF-12 physical well-being score.
The authors note that “the longer-term implications of these findings are unclear,” but their calculations indicated that, through 20 years, none of the XLPE liners would wear through, but 6 of the conventional PE liners would require revision due to wear-through.